TaqMan™ Array Human MicroRNA A Cards v2.0
TaqMan™ Array Human MicroRNA A Cards v2.0
Applied Biosystems™

TaqMan™ Array Human MicroRNA A Cards v2.0

The TaqMan® Array Human MicroRNA A Card v2.0 contains 384 TaqMan® MicroRNA Assays enabling accurate quantitation of 377 human microRNAs詳細を見る
製品番号(カタログ番号)数量
43989654 pack
製品番号(カタログ番号) 4398965
価格(JPY)
330,900
Each
お問い合わせください ›
数量:
4 pack
The TaqMan® Array Human MicroRNA A Card v2.0 contains 384 TaqMan® MicroRNA Assays enabling accurate quantitation of 377 human microRNAs in as little as 5 hours. Three TaqMan MicroRNA Assay endogenous controls are included to aid in data normalization and one TaqMan® MicroRNA Assay not related to human is included as a negative control. Details on the included assays is available in the Target List file. Purchase of the Megaplex™ RT Primers, Human Pool A v2.1 is required to run this array. An additional preamplification step is enabled by Megaplex™ PreAmp Primers, Human Pool A v2.1 for situations where sensitivity is of the utmost importance and⁄or sample is limiting.
研究用途にのみご使用ください。診断目的には使用できません。
仕様
使用対象 (装置)AB 7900 HT 384-Well System
形状乾燥
内部プローブの変更MGB (Minor Groove Binder) (3'), Non-Fluorescent Quencher (3'), FAM (5')
標識または色素FAM
製品ラインMegaplex, TaqMan
数量4 pack
反応速度Standard
研究カテゴリーGene Regulation, Epigenetics
タイプArray Human MicroRNA A Card
フォーマット384-well Microfluidic Card
Human
Unit SizeEach
組成および保存条件
4°C

よくあるご質問(FAQ)

Why does the negative control well show amplification when doing microRNA analysis using Megaplex Primer Pools and TaqMan Array Cards?

Most likely the reagents or the cDNA template are contaminated. Please follow established PCR laboratory best practices.

Why do I have poor reproducibility across technical replicates when doing microRNA analysis using Megaplex Primer Pools and TaqMan Array Cards?

Most likely the reagents were not adequately mixed. Ensure that all samples and reagents are mixed well.

When doing microRNA analysis using Megaplex Primer Pools and TaqMan Array Cards, why are the Ct values for the endogenous controls highly variable across the sample set?

Highly variable values for endogenous controls is most likely due to biological variation. We recommend that you use an alternative endogenous control, such as a non-variable miRNA.

Why is the Ct value for the no-template control (NTC) <35 for some TaqMan microRNA Assays?

The Ct value for the NTC can be <35 for the following reasons:
- There are non-specific interactions between primers. For NTC information on a specific assay, see the Megaplex Assay Performance File.
- The cDNA template is contaminated. Please ollow established PCR good laboratory practices to prevent contamination.
- The preamplification product was not diluted properly before real-time PCR. Ensure that you dilute the preamplification product according the procedure in the user guide.

Why can I not detect any miRNA in my sample using the TaqMan MicroRNA Assay?

Most likely, the threshold is set too high to detect miRNA in samples with low levels of expression. We recommend adjusting the threshold (Ct) or NOAMP flag threshold (Crt) to an appropriate level.