Retainer & Base Set (Standard) for 3500xL Genetic Analyzer, 384 well
Product Image
Applied Biosystems™

Retainer & Base Set (Standard) for 3500xL Genetic Analyzer, 384 well

This 384 -well plate Retainer and Base Standard Set is for use with the 3500xL Genetic Analyzer.For Research Use Only.詳細を見る
製品番号(カタログ番号)数量
44102354 Sets
製品番号(カタログ番号) 4410235
価格(JPY)
31,300
Each
お問い合わせください ›
数量:
4 Sets
This 384 -well plate Retainer and Base Standard Set is for use with the 3500xL Genetic Analyzer.

For Research Use Only. Not for use in diagnostics procedures.
研究用途にのみご使用ください。診断目的には使用できません。
仕様
使用対象(アプリケーション)配列
使用対象 (装置)3500xL Genetic Analyzer
製品タイプBase Set
数量4 Sets
出荷条件室温
フォーマット384-well Plate
Unit SizeEach
組成および保存条件
This product includes 4 sets of retainers and bases.

よくあるご質問(FAQ)

When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?

Some of the other causes of an Unstable Electrophoresis current detected error message are:

1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues

Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.

Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?

If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.

Find additional tips, troubleshooting help, and resources within our Capillary Electrophoresis Instruments Support Center.