This is a cap for use with POP-4™, POP-6™, and POP-7™ polymer pouches.
For Research Use Only. Not for use in diagnostics procedures.
研究用途にのみご使用ください。診断目的には使用できません。
仕様
使用対象(アプリケーション)配列
使用対象 (装置)3500 Series Genetic Analyzers, SeqStudio Flex Series Genetic Analyzers
製品タイプPouch Cap for Polymer Pouch
数量4 Pieces
出荷条件室温
フォーマットPouch
Unit SizeEach
組成および保存条件
Each package contains 4 caps.
よくあるご質問(FAQ)
When I started my run on the Applied Biosystems 3500/3500xL Genetic Analyzer, an error occurred: "Unstable Electrophoresis current detected, check for air bubbles" but I don't see any bubbles. What else can cause this?
Some of the other causes of an Unstable Electrophoresis current detected error message are:
1.Leak on the system
2. Polymer that:
-Has expired
-Has been left on the instrument for more than the recommended time
-Is a mixture of expired polymer and non-expired polymer
3. Running Buffer that was:
-On the instrument longer than 2 weeks
-Not filled to the fill line or evaporated below the fill line
4. An arcing event that was not cleaned afterwards using the water wash wizard
5. Not performing regular maintenance on the instrument
6. Hardware issues
Inspect the system for leaks. If you do not see any leaks on the system, perform the wash the pump chamber and channels wizard using the conditioning pouch and place fresh, non-expired polymer and fresh Anode and Cathode buffer containers. If the problem persists, a service call may be required.
Do I need to replace the buffer every 14 days if the Applied Biosystems 3500/3500xL Genetic Analyzer is not in use?
If the instrument is not in use, it is not necessary to replace the buffer every 14 days. However, the buffer level needs to remain at the fill line to prevent the capillary or array from drying out. Top off the buffer volume with water if necessary. Replace the anode and cathode buffer containers prior to starting a new run.