CaptureSelect™ KappaXP MiniChrom™ Pre-packed columns
Thermo Scientific™

CaptureSelect™ KappaXP MiniChrom™ Pre-packed columns

The Thermo Scientific™ CaptureSelect™ KappaXP MiniChrom™ column is designed for bench-scale resin screening, process development, and sample preparation.
製品番号(カタログ番号)数量
59432120055 mL
製品番号(カタログ番号) 5943212005
価格(JPY)
260,700
Each
数量:
5 mL

CaptureSelect KappaXP MiniChrom™ column is designed for bench-scale resin screening, process development, and sample preparation.

This pre-packed column is ready to connect and has the following features:

  • Volume of 5 mL (0.8 x 10 cm)
  • Direct connection to standard chromatography systems
  • Packed with CaptureSelect KappaXP Affinity Matrix, offering high dynamic binding capacity and high resolution over a wide range of flow rates

The 10-cm bed height of the 5-mL column allows initial process development on a bench scale.

CaptureSelect KappaXP matrix recognizes the kappa light chain of human antibodies, enabling purification of all Kappa containing Ig's, Fab, and Fab2 fragments directly from complex sources and feed streams.

Features of this affinity matrix include:

  • Purification of Kappa monoclonal antibodies, Fab, and Fab2 fragments in a single step with high purity and yield
  • Mild elution for Fab fragments and antibodies
  • Human-specific, no binding to bovine antibodies
  • Excellent scalability
  • Excellent pH stability up to 100 mM NaOH (limited exposure)
  • Non-animal derived

Free of animal components

CaptureSelect products contain affinity ligands based on recombinant single-domain antibody fragments (VHH) created by a proprietary technology. The VHH affinity ligand is a 12 to 15 kDa fragment comprising the three complementarity-determining regions (CDRs) that form the antigen-binding domain, efficiently produced in the yeast Saccharomyces cerevisiae by a production process free of any animal components (Animal Origin-Free).

Main characteristics

  • Matrix: agarose-based, epoxide-activated
  • Average particle size: 65 ± 10 μm
  • Ligand: CaptureSelect KappaXP affinity ligand
  • Ligand coupling method: epoxide coupling
  • Fab binding capacity: 20–30 g IgG Kappa-Fab/liter resin depending on flow rate, column height, and residence time
  • IgG binding capacity: 35–45 g IgG/liter resin depending on flow rate, column height
  • Elution conditions: 20 mM citric or acetic acid, pH 3–4; 100 mM Tris, 1.5 M MgCl2, pH 6
  • Column dimensions: 5 mL (0.8 x 10 cm)
  • Flow characteristics: 75–150 cm/h (up to 2 bar)
  • Formulation buffer: 20% (v/v) ethanol
研究用途にのみご使用ください。診断目的には使用できません。
仕様
ベッド容量5 mL
カラムタイプMiniChrom
直径(メートル法)0.8 cm
流速75 to 150 cm/hr. (Up to 2 bar)
フォーマットChromatography Cartridge
長さ(メートル法)10 cm
リガンドタイプKappaXP affinity ligand
数量5 mL
出荷条件Wet Ice
固定相KappaXP Affinity Matrix with KappaXP Ligand
粒子径65 ±10 μm
製品ラインCaptureSelect
タイプPre-packed Column
Unit SizeEach
組成および保存条件
Store in refrigerator (2°C to 8°C).