Thermo Scientific™

Direct Methane & Non-Methane Hydrocarbon Analyzer Model 55i

Catalog number: 55I
Thermo Scientific™

Direct Methane & Non-Methane Hydrocarbon Analyzer Model 55i

Catalog number: 55I
Make automated measurements of methane and non-methane hydrocarbons with the Thermo Scientific™ Direct Methane & Non-Methane Hydrocarbon Analyzer, Model 55i. Unlike instruments that measure only methane and total hydrocarbons, this back-flush gas chromatography (GC) system provides a direct measurement of non-methane concentrations.
 
Catalog Number
55I
Unit Size
Each
Description
Direct Methane and Non-Methane Hydrocarbon Analyzer
Price (USD)
Full specifications
Detectable AnalytesMethane and Nonmethane Hydrocarbons
Flow Rate0.5 L/min.
Inputs16 Digital Inputs, User Select Programmable, TTL Level, Pulled High
Linearity±2% of span (at concentrations of 10 to 100% of span)
Response Time70 sec.
Span Drift<2% span value (without auto calibration)
Zero Noise0.025 ppm RMS
Accuracy±1% of range
Certifications/ComplianceETL, CSA, UL, CE, UKCA
Depth (English)23 in.
Depth (Metric)584 mm
DescriptionDirect Methane and Non-Methane Hydrocarbon Analyzer
Detection Limit0.050 ppm CH4 - 0.050 NMHC as propane (300 second averaging time)
Frequency50/60 Hz
Height (English)8.62 in.
Height (Metric)219 mm
Measurement Range0 to 5, 50, 500 ppm, or 0 to 10, 100, 1000 ppm, or 0 to 20, 200, 2000 ppm , or 0 to 50, 500, 5000 ppm (Group selectable at time of order)
Mounting OptionsBench, EIA rack, Retrofit rack
OutputsAnalog outputs: 6 voltage outputs; 0 to 100mV, 1, 5, 10V (user selectable), 5% of full-scale over/under range (user selectable), 12 bit resolution, measurement output user selectable per channelDigital outputs: 1 power fail relay Form C, 10 relays Form A, user-selectable alarm output, relay logic, 100mA at 200VDCDigital inputs: 16 digital inputs, user select programmable, TTL level, pulled high
Precision2.0% of reading or 50 ppb (whichever is larger)
Voltage100/115/220/240 V
Weight (English)60 lb.
Weight (Metric)17.7 kg
Width (English)16.75 in.
Width (Metric)425 mm
Unit SizeEach
Showing 1 of 1
Catalog NumberSpecificationsUnit SizeDescriptionPrice (USD)
55IFull specifications
EachDirect Methane and Non-Methane Hydrocarbon AnalyzerRequest A Quote
Detectable AnalytesMethane and Nonmethane Hydrocarbons
Flow Rate0.5 L/min.
Inputs16 Digital Inputs, User Select Programmable, TTL Level, Pulled High
Linearity±2% of span (at concentrations of 10 to 100% of span)
Response Time70 sec.
Span Drift<2% span value (without auto calibration)
Zero Noise0.025 ppm RMS
Accuracy±1% of range
Certifications/ComplianceETL, CSA, UL, CE, UKCA
Depth (English)23 in.
Depth (Metric)584 mm
DescriptionDirect Methane and Non-Methane Hydrocarbon Analyzer
Detection Limit0.050 ppm CH4 - 0.050 NMHC as propane (300 second averaging time)
Frequency50/60 Hz
Height (English)8.62 in.
Height (Metric)219 mm
Measurement Range0 to 5, 50, 500 ppm, or 0 to 10, 100, 1000 ppm, or 0 to 20, 200, 2000 ppm , or 0 to 50, 500, 5000 ppm (Group selectable at time of order)
Mounting OptionsBench, EIA rack, Retrofit rack
OutputsAnalog outputs: 6 voltage outputs; 0 to 100mV, 1, 5, 10V (user selectable), 5% of full-scale over/under range (user selectable), 12 bit resolution, measurement output user selectable per channelDigital outputs: 1 power fail relay Form C, 10 relays Form A, user-selectable alarm output, relay logic, 100mA at 200VDCDigital inputs: 16 digital inputs, user select programmable, TTL level, pulled high
Precision2.0% of reading or 50 ppb (whichever is larger)
Voltage100/115/220/240 V
Weight (English)60 lb.
Weight (Metric)17.7 kg
Width (English)16.75 in.
Width (Metric)425 mm
Unit SizeEach
Showing 1 of 1

Automated batch analyzer repeatedly collects and analyzes small amounts of the sample stream drawn in by the pump. Central to the instrument's operation is an eight-port, two-position rotary valve which is used to introduce the gas sample into the analyzer and to control the flow of gases through the chromatographic column. The proprietary column design is unaffected by the oxygen content of the sample and provides complete recovery of low volatility compounds while achieving absolute separation of methane from all C2 compounds.

  • Adjustable ranges
  • Real-time correction of THC readings
  • Automatic flame sensing and ignition
  • Automatic calibration and span check
  • 70-second analysis
  • In-line particulate filter

iSeries Advantage:

  • Ethernet connectivity for efficient remote access
  • Enhanced user interface with one button programming and large display screen
  • Flash memory for increased data storage and user downloadable software
  • Enhanced electronics design optimizes product commonality
  • Improved layout for easier accessibility to components

Figures

Documents & Downloads

Certificates

    Frequently asked questions (FAQs)

    Citations & References