Specifications
Power Consumption180 VA
Time and Temperature ProfilesPreheat: LiF:Mg, Ti:0 sec., 50°C, LiF:Mg, Cu, P:10 sec., 165°C; Aquisition: LiF:Mg, Ti:13.33 sec; 25 °C/sec., 300°C, LiF:Mg, Cu, P:13.33 sec.; 15°C, 260°C
Capacity200 personnel/environmental, 400 EXTRAD or 800 DXTRAD extremity dosimeters per loading
Card IDCodabar, interlaced 2 of 5, code 39
Dark Current<1μGy 137Cs equivalent dark current
DescriptionHarshaw TLD Model 6600 Plus Automated Reader Instrument
Dose RangeLiF:Mg, Ti: 10μGy to 1 Gy, linear within 5%;1 Gy to 100 Gy, supralinear above 1 Gy
LiF:Mg, Cu, P:1μGy to 20 Gy, linear within 5%; 20 Gy to 100 Gy, sublinear above 20 Gy
Flow RateDry prepurified N2 gas or dry air
Idle Mode, 28 L/hr. (1 SCFH)
Maximum, 850 L/hr. (30 SCFH)
Heating MethodLinear controlled contactless heating using N2 gas or dry air
Hertz100, 200, 220, 240 V ±10%, 50/6 Hz
LinearityLiF:Mg, Ti; <5%, 10μGy to 1 Gy
Radiation TypePhotons, Neutrons, Beta
Reference Light14C-activated Ca2F:Eu
Repeatability<2% variation based on one standard deviation for 10 repeated doses measurements of 1 mGy (100 mrad) from 137 Cs
StabilityBased on one standard deviation of 10 consecutive measurements: Dose, better than 1μGy; Reference light, short-term stability, <0.5%; High voltage, short-term stability, ±0.005
Operating Temperature (English)32°F to 104°F
Storage Temperature (English)14°F to 140°F
Operating Temperature (Metric)0°C to 40°C
Storage Temperature (Metric)-10°C to +60°C
Throughput(LiF:Mg, Ti standard heating); 4-element dosimeters, 70/hr; 2-element dosimeters, 120/hr; EXTRAD dosimeters, 240/hr; DXTRAD dosimeters, 280/hr
Warm Up Time<20 min.
Weight (English)154 lb.
Weight (Metric)70 kg
Product TypeRadiation Dosimeter Reader
Detector TypeThermoluminescence (TL) Detector with LiF material
SensitivityDependent on config. Used
Unit SizeEach