Combine the precision of light-scattering nephelometry with the accuracy of beta attenuation technology with the Thermo Scientific™ Model 5030i SHARP (Synchronized Hybrid Ambient Real-time) particulate monitor. This automated, continuous, non-step wise, ambient particulate monitor measures PM-10 and PM-2.5 and is built on the proven iSeries design platform.

 
Catalog Number
5030I
Unit Size
Each
Description
Model 5030i SHARP Monitor
Price (USD)
Full specifications
Flow Rate14 to 20 L/min.
Inputs16 Digital Inputs (Standard), 8 0 to 10 VDC Analog Inputs (Optional) 8 User-Defined Analog Outputs (0 -1 or 0 -5 VDC)
Span Drift0.02% (24 hr.)
Temperature (English) Operating39.2°F to 122°F
Temperature (Metric) Operating4°C to 50°C
Accuracy5% using NIST-traceable mass foil set
Certifications/ComplianceDesigned to meet CE: EN61326: 1997 + A1: 1998 + A2: 2001 + A3: 2003, EN61010-1UL: 61010-1: 2004; CSA: C22.2 No. 61010-1: 2004; FCC: Part 15 Subpart B, Class B
Depth (English)23 in.
Depth (Metric)58.4 cm
DescriptionModel 5030i SHARP Monitor
Detection Range6µg/m³, (0.5hour), 4µg/m³ (24-hour) ; 3µg/m³ (3-hour), 1µg/m³ (24-hour)
Electrical Requirements100-240 VAC, 50-60 Hz recommended, 805 watts (115V); 880 watts maximum (220-240 V (instrument, heater & pump). Pump: 120 VAC/60 Hz; 4.25A; 240 VAC/ 50Hz; 2.25 A
Height (English)16 in.
Height (Metric)36 cm
Measurement Range0 to 1.0, 2.0, 3.0, 5.0, 10.0 mg/m3; 0-100, 1,000, 2,000, 3,000, 5,000, 10,000 μg/m3
OutputsRS232/RS485, TCP/IP, 10 status relays and power fail indication (standard). 9 user defined analog outputs (0-100mV, 0-1, 0-5 or 0-10 VDC), six 0-20 or 4-20mA isolated current output (optional)
Precision±2.0 μg/m3 <80 μg/m3, 4 to 5 μg/m3 >80 μg/m3 (24-hr.)
Resolution0.1μg/m3 , updated every 1 second
Weight (English)57.6 lb.
Weight (Metric)26.1 kg
Width (English)6.75 in.
Width (Metric)42.5 cm
Unit SizeEach
Showing 1 of 1
Catalog NumberSpecificationsUnit SizeDescriptionPrice (USD)
5030IFull specifications
EachModel 5030i SHARP MonitorRequest A Quote
Flow Rate14 to 20 L/min.
Inputs16 Digital Inputs (Standard), 8 0 to 10 VDC Analog Inputs (Optional) 8 User-Defined Analog Outputs (0 -1 or 0 -5 VDC)
Span Drift0.02% (24 hr.)
Temperature (English) Operating39.2°F to 122°F
Temperature (Metric) Operating4°C to 50°C
Accuracy5% using NIST-traceable mass foil set
Certifications/ComplianceDesigned to meet CE: EN61326: 1997 + A1: 1998 + A2: 2001 + A3: 2003, EN61010-1UL: 61010-1: 2004; CSA: C22.2 No. 61010-1: 2004; FCC: Part 15 Subpart B, Class B
Depth (English)23 in.
Depth (Metric)58.4 cm
DescriptionModel 5030i SHARP Monitor
Detection Range6µg/m³, (0.5hour), 4µg/m³ (24-hour) ; 3µg/m³ (3-hour), 1µg/m³ (24-hour)
Electrical Requirements100-240 VAC, 50-60 Hz recommended, 805 watts (115V); 880 watts maximum (220-240 V (instrument, heater & pump). Pump: 120 VAC/60 Hz; 4.25A; 240 VAC/ 50Hz; 2.25 A
Height (English)16 in.
Height (Metric)36 cm
Measurement Range0 to 1.0, 2.0, 3.0, 5.0, 10.0 mg/m3; 0-100, 1,000, 2,000, 3,000, 5,000, 10,000 μg/m3
OutputsRS232/RS485, TCP/IP, 10 status relays and power fail indication (standard). 9 user defined analog outputs (0-100mV, 0-1, 0-5 or 0-10 VDC), six 0-20 or 4-20mA isolated current output (optional)
Precision±2.0 μg/m3 <80 μg/m3, 4 to 5 μg/m3 >80 μg/m3 (24-hr.)
Resolution0.1μg/m3 , updated every 1 second
Weight (English)57.6 lb.
Weight (Metric)26.1 kg
Width (English)6.75 in.
Width (Metric)42.5 cm
Unit SizeEach
Showing 1 of 1
The Model 5030i SHARP monitor is developed on the renowned Thermo Scientific iSeries platform design and offers unprecedented ease of use. The monitor's light-scattering nephelometry is checked against the mass concentration of the beta attenuation data for calibration. It utilizes non-stepwise measurement to auto-advance the particulate-laden sample, avoiding potential particle loss.

Key Features:

  • High time resolution, one-minute measurements
  • Real-time data
  • Digital dynamic filtering for continuous calibration
  • Superior accuracy, precision and sensitivity
  • Continuous, non-stepwise measurement provides long-term unattended operation
  • Volatile loss mitigation via Dynamic Heating System
  • Enhanced user interface

Additional Key Features:

  • Flash Memory for increased data storage
  • Enhanced Ethernet connectivity
  • Remote data access
  • Short-cut keys with easy programming
  • Large, easy-to-read interface screen
  • iPort communication software
  • Modular design for future upgrades and enhancements

Figures

Documents & Downloads

Certificates

    Frequently asked questions (FAQs)

    Citations & References

    Search citations by name, author, journal title or abstract text