A compact system with a stainless-steel frame that holds all components, Thermo Scientific™ Reverse Osmosis EDI features a microprocessor control for fully automatic control and monitoring of the system. Limiting value setting and quality rinse are standard.

 
Catalog Number
07.3900-EDI
Unit Size
Each
Permeate performance
800 L/hr.
Price (USD)
Full specifications
Bacterial Content99%
Depth (Metric)600 mm
DescriptionReverse Osmosis EDI 800 System
Electrical Requirements400 V 50 Hz
Frequency50 Hz
Height (Metric)1800 mm
Net Weight (Metric)360 kg
Operating Pressure14 bar
Permeate performance800 L/hr.
Voltage400 V
Width (Metric)1400 mm
Capacity (Metric)300 to 1100 L/hr
Feed Water MonitoringYes
Flow Rate800 L/hr.
Product Water TypeReverse Osmosis (Type 3)
TypeReverse Osmosis System
Unit SizeEach
Showing 1 of 1
Catalog NumberSpecificationsUnit SizePermeate performancePrice (USD)
07.3900-EDIFull specifications
Each800 L/hr.Request A Quote
Bacterial Content99%
Depth (Metric)600 mm
DescriptionReverse Osmosis EDI 800 System
Electrical Requirements400 V 50 Hz
Frequency50 Hz
Height (Metric)1800 mm
Net Weight (Metric)360 kg
Operating Pressure14 bar
Permeate performance800 L/hr.
Voltage400 V
Width (Metric)1400 mm
Capacity (Metric)300 to 1100 L/hr
Feed Water MonitoringYes
Flow Rate800 L/hr.
Product Water TypeReverse Osmosis (Type 3)
TypeReverse Osmosis System
Unit SizeEach
Showing 1 of 1

Technical benefits at a glance

  • Capacity is 500 -1100L/hr. and is expandable
  • At least 75% yield of high purity water
  • 99% reduction in salts, bacteria, particles and pyrogens
  • Automatic rinsing cycles
  • Low cost doubling of capacity on increased need for high purity water
  • Feed water is softened potable water
  • Housing is a stainless steel frame
  • System is not to be wall mounted
  • Pretreatment - carbon prefilter
  • Digital conductivity of feed and permeate
  • Display shows errors
  • Flow meter included
  • RS232 optional
  • Storage tank is not included but required

High purity water for exceptional cases

The quality of the high purity water supplied by a reverse osmosis membrane is not sufficient when a conductivity below 1μS/cm is required. For such cases, a combination of reverse osmosis with downstream electrodionization (EDI) is needed. Results from routine usage show that such a combination with EDI technology allows product water of less than 0.1μS/cm to be obtained from 600μS/cm raw water. In addition, the continual self-regeneration of the high purity EDI mixed bed resins effects a very high microbiological purity of the water produced.

Efficient performance, economical running

  • For supplying autoclaves, washing machines and automatic analyzers when the tap water has a high salt content
  • When ion exchanger regeneration costs are high
  • When reliable handling of the regeneration process cannot be ensured
  • When greater-than150L of high purity water are required daily
  • When removal of a high percentage of bacteria, particle and pyrogen is required
  • When a constant water quality is a necessity
  • As central purified water system in buildings

Dependable high purity water quality

Thermo Scientific Barnstead reverse osmosis (RO) membranes are not only extremely effective in removing salts, but also more than 99% effective in removing incoming bacteria, particles and pyrogens.

The quality rinse is a completely automatic rinsing cycle with permeate discard before and after use, as well as when not in use, to ensure constant high purity water quality.

More water needed? No problem.

Should your high purity water requirement increase (e.g. because of the purchase of additional glassware washers), then your system can be retrofitted to double its capacity without changing its dimensions!

Saves space and cuts costs

A central water treatment system replaces numerous individual systems and so not only saves space but also eliminates costs for ion exchanger regeneration etc.

Compact but with room for more

Ready-to-connect systems are available for mounting on a wall. Integrated pretreatment reduces loss in performance, e.g. due to suspended matter and precipitates of mineral salts or hardness formers, can easily be integrated inside the modular system cabinet.

Downstream treatment measures, e.g. sterile filtration, UV disinfection and tanks, can also be mostly fitted in, according to their size.

Ordering Information:

This product is not available in all regions

Figures

Frequently asked questions (FAQs)

Citations & References