Applied Biosystems™

TaqMan™ RNase P Instrument Verification Plate for 7900HT System, 384-well

Catalog number:  4323306

Related applications:

Real Time PCR (qPCR)

Back to top

Description

The Applied Biosystems® TaqMan® RNase P 384-Well Instrument Verification Plate is used to verify the performance of the Applied Biosystems® 7900HT Fast Real-Time PCR System. The TaqMan® RNase P 384-Well Instrument Verification Plate is a 384-well PCR plate preloaded with the necessary reagents to detect and quantitate genomic copies of the human RNase P gene, a single-copy gene encoding the RNA moiety of the RNase P enzyme. Each well contains preloaded reaction mix (1X TaqMan® Universal PCR Master Mix, RNase P primers, and FAM™ dye–labeled probe) and template.
The TaqMan® RNase P Instrument 384-Well Instrument Verification Plate satisfies the same requirements as the Applied Biosystems® 7900 HT Real-Time PCR Systems installation. To pass installation, the instruments must demonstrate the ability to distinguish between 5,000 and 10,000 genomic equivalents with a 99.7% confidence level for a subsequent sample run in a single well. To ensure accurate results, Applied Biosystems® recommends that the performance of your instrument be verified every six months.
For Research Use Only. Not for use in diagnostic procedures.

Specifications

Detection Method: Fluorescent
Label or Dye: FAM
Product Line: TaqMan™
Shipping Condition: Dry Ice
Calibration Type: RNase P Calibration, Instrument Verification Plate
Description: For 384-well
For Use With (Equipment): AB 7900 HT 384-Well System
Includes: Contains a 384-well Reaction Plate pre-loaded and sealed with the 5’ nuclease assay using RNase P TaqMan™ probes and primers.
Quantity: 1 x 384-well Plate

Contents & storage

Contains a 384-Well Reaction Plate pre-loaded and sealed with the 5’ nuclease assay using RNase P TaqMan® probes and primers. Store at -15 to 25°C. Protect from light.