Antaris™ MX FT-NIR Process Analyzer
Antaris™ MX FT-NIR Process Analyzer
Thermo Scientific™

Antaris™ MX FT-NIR Process Analyzer

Optimize your QC testing and remotely monitor the process using fiber optic simultaneous multiplexing technology with the Antaris™ MX FT-NIR process analyzer.

Have Questions?
Catalog number IQLAADGABCFADUMAAZ
Price (USD)/ Each
-
Request A Quote
Description:
Antaris MX FT-NIR Process Analyzer
Analyzer Type:
Fourier Transform Near-Infrared
Source Type:
Long-life, high-intensity halogen NIR source; spare source included with system, guaranteed filament image alignment
Spectral Range:
12,000 to 3800 cmto1 (833 to 2500 nm)

Optimize your QC testing while remotely monitoring the process using fiber optic simultaneous multiplexing technology and integrated communication for real-time feedback with the Thermo Scientific™ Antaris™ MX FT-NIR Process Analyzer. The Antaris MX FT-NIR Process Analyzer provides a complete, fit-for-purpose solution for point-of-use materials analysis and online process monitoring applications.

This easy-to-use NIR analyzer allows a single process to be monitored at multiple points or multiple processes to be monitored simultaneously. The Antaris MX Process Analyzer also eliminates removing the probe from the process stream for background collection or qualification.

The Antaris MX FT-NIR Process Analyzer Features:

  • Thermo Scientific™ ValPro™ Instrument Qualification Package
  • Always-on internal background
  • SMA905 industry-standard connectors for fiber optics — choose the probe right for your process
  • OPC, 4-20mA and Digital I/O for process communications

Ordering Information:

The Antaris MX FT-NIR process analyzer is sold as a configured system designed to meet your specific application needs. For more information, contact your Thermo Scientific Sales Representative.

Recommended for:

  • In-process liquid analysis
  • In-process solid analysis
  • Laboratory liquid analysis
  • Laboratory solid analysis
Specifications
DescriptionAntaris MX FT-NIR Process Analyzer
Analyzer TypeFourier Transform Near-Infrared
DetectionHigh sensitivity, high stability matched InGaAs
Height (English)18.97 x 15.98 x 9.72 in.
Height (Metric)48.2 x 40.6 x 24.7 cm
InterfaceDirect PC to ethernet allows file system and OPC communications
InterferometerProven, frictionless, stable, long-life Michelson
Operating Temperature Range15°C to 35°C (59° to 95°F)
Photometric LinearitySlope 1.0 ±0.05 and an intercept of 0.0 ±0.05
Resolution4 cm-1 across spectral range (0.6 nm at 1250 nm), 2 cm-1 option across spectral range (0.3 nm at 1250 nm)
Sampling ModeFiber optic multiplexing
Sealed and DesiccatedYes
Source TypeLong-life, high-intensity halogen NIR source; spare source included with system, guaranteed filament image alignment
Spectral Range12,000 to 3800 cmto1 (833 to 2500 nm)
System Status IndicatorsIndicator lights report scan, laser, power, and source status continuously
Voltage90/264 V
Wavenumber Accuracy±0.1 cm-1 (0.02 nm at 1250 nm)
Wavenumber Repeatability(System to System) Better than 0.05 cm-1 (0.008 at 1250 nm)
Wavenumber Reproducibility(Single System) 10 measurement standard deviation <0.006cm-1
Weight (English)105.16 lb.
Weight (Metric)25.8 kg
Unit SizeEach